5-4
Making Measurements
To use delayed sweep
To use delayed sweep
You can use the delayed sweep window to locate and horizontally expand part
of the main sweep for a more detailed (high-resolution) analysis of signals.
1 Connect a signal to the oscilloscope and obtain a stable display.
2Press the
Main/Delayed key.
3Press the
Delayed softkey.
Delayed sweep display
Time/div for
main sweep
Time/div for
delayed sweep
Delay time momentarily
displays when the delay
time knob is turned
Set trigger point time
reference to left, right
or center
Select main or
delayed sweep
Delayed sweep
window
Main sweep
window
These markers define the
beginning and end of the
delayed sweep window