Chapter 6: Replacing Assemblies
To remove and replace the CD-ROM drive
128
To remove and replace the CD-ROM drive
Use this procedure to remove and replace the CD-ROM drive. When necessary, refer to other
removal procedures.The graphics in this chapter are representative of the oscilloscope at the
time of this printing. Your unit may look different.
1
Disconnect the power cable from the instrument and remove the top sleeve.
2 Remove the following cables:
SATA Cable
CD-ROM Power cable
3
Using a Torx T10 driver, remove the three CD-ROM support bracket screws holding
the CD-ROM drive in place.
When re-assembling, torque the three Torx T10 screws to 5 in-lb.
4
Using a T20 driver remove the rear panel screw holding the CD-ROM drive in place.
When re-assembling, torque the Torx T20 screw to 18 in-lb.
5
Move the CD-ROM drive back towards the front frame, then lift and pull to remove.
Figure 6-26
Removing the CD-ROM assembly
Torx
T20
SATA cable
Power
cable
Torx
T10
Torx
T10
Torx
T10

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infiniium 80000 Series and is the answer not in the manual?

Agilent Technologies Infiniium 80000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniium 80000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals