Performance Tests 7
138 Agilent U1602B/U1604B User’s and Service Guide
Bandwidth Verification Test
This test verifies bandwidth of the instrument. In this test, you will use an
oscilloscope calibrator to output a sine wave signal at 1 MHz and
bandwidth frequency (20 MHz and 40 MHz). You will use the peak- to- peak
voltage to verify the bandwidth response of the instrument.
U1602B
Test limits ¡ –3 dB at 20 MHz (for all channels)
U1604B
Test limits ¡ –3 dB at 40 MHz (for all channels)
Refer to Table 7- 1 for the equipment required for this bandwidth
verification test.
1 Connect the calibrator output through a 50
Ω feedthrough termination
to input channel 1 of the instrument.
2 Set up the instrument.
Set the time base to 500 ns/div.
Set the Volts/Div for channel 1 to 200 mV/div.
Select User mode of the instrument, then press F2 softkey for
Acquire mode.
Press F1 to select Average acquisition mode. Turn the rotary switch
to select averaging number of 8 (Avg 8). Press the rotary switch to
set the setting.
3 Set the calibrator to "Level Sine" and OPR/STBY to "OPR". Set the
calibrator to 1 MHz and six divisions of amplitude.
NOTE
The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m), 1
meter long and Fluke P/N 686318.

Table of Contents

Other manuals for Agilent Technologies U1602B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1602B and is the answer not in the manual?

Agilent Technologies U1602B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1602B
CategoryTest Equipment
LanguageEnglish

Related product manuals