5. Advanced operation
• Native:IfyouselectNative, the native panel behavior will not be corrected.
• sRGB:Thisisthedisplayfunctionasdefined in the sRGB specificationandisdesignedtomatch
typical home and office viewing conditions. It is widely used in most computer applications.
• DICOM: DICOM (Digital Imaging and Communications in Medicine) is an international standard that
was developed to improve the quality and communication of digital images in radiology. In short,
the DICOM gamma function results in more visible grayscales in the images. Barco recommends
selecting the DICOM gamma function for most medical viewing applications.
• Gamma 2.2: Select this function in case the display is to replace a CRT display with a gamma of 2.2.
• QAWeb: This gamma function will be automatically selected when gamma functions are defined by
MediCal QAWeb.
To select a gamma function
1. Bring up the OSD main menu.
2. Navigate to the Adjustments menu.
3. Enter the Gamma submenu.
4. Select one of the available gamma functions.
5.4 Ambie nt light reading room selection
About ambient light reading rooms
The available ambient light reading rooms for your display are:
• Dark Room: Corresponds to light conditions in dark diagnostic reading rooms. This setting has the
lowest maximum ambient light.
• Office: Corresponds to light conditions in office rooms.
• Operation Room: Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
• QAWeb: This setting will be automatically selected when ambient li
ght conditions are defined by Med-
iCal QAWeb.
To select an ambient light reading room
1. Bring up the OSD main menu.
2. Navigate to the Adjustments menu.
3. Enter the Ambient Light submenu.
4. Select one of the available reading rooms and confirm.
5.5 W hite p oin t selection
About white point selection
This setting allows you to modify the display white point, used as reference for all other colors to be dis-
played.
The available white point settings for your display are:
• Native: The native, unmodified color temperature of the LCD panel.
• 6500K (sRGB): Corresponds to a color temperature of 6500 Kelvin (D65).
• QAWeb: This white point setting will be automatically selected when white point is defined by
MediCal QAWeb.
16