G2S A20 / G2S C20 User Manual V1.0
24 / 39
14.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Range
Description
Humidifier
Off, 1 5
There are five humidity levels available. As the
numbering increases, the humidity rises
accordingly. “Off” means the humidifier is turned
off. The default setting is 2
Tube
15mm / 22mm
Setting diameter of the tube by adjusting this
option. The default setting is “22mm
Reslex
Off, 1 3
This feature enables the device to automatically
reduce the treatment pressure when the patient
exhales, so as to make the user more comfortable.
The higher the numbering is, the more pressure
the device reduces. “Off” means this feature is
disabled. The default setting is Off
Ramp Time
0 - Max Ramp
In order to increase comfort and help the patient
fall asleep easily, the pressure can increase
gradually, when the Ramp feature is enabled. The
ramp time during which the initial pressure rises to
the prescribed treatment pressure can be
adjusted. As you turn the Knob to the
nearest point, the numbering increases or
decreases by five minutes. The default setting is
10 minutes. The screen displays a real-time
countdown of the remaining ramp time in minutes
Delay
On / Off
When the humidifier is on, this feature allows the
airflow to continue for about 15 minutes at a low
pressure (about 2 hPa) after you press the Knob
to discontinue treatment. This will blow off the
vapor left in the water chamber to avoid any
damage to the device. When this feature is set to
“Off,” which means it is disabled, the airfolw stops
delivering air instantly after you press the Knob
. The default setting is “Off
Date
2000-01-01
2099-12-31
Set date by adjusting this option
Time
00:00
23:59
Set time by adjusting this option
Brightness
High / Low
Setting screen brightness by adjusting this option.
The default setting is “High
Mask Type
Full Face;
Nasal; Pillow;
Other
There are three mask types available, Full Face
(full-face mask), Nasal (nasal mask), and Pillow
(nasal pillow mask). The default mask type is
Nasal, but the patient can choose other suitable
masks as well. When selecting masks other than
the above three types of BMC masks, the patient
can identify the masks as other.

Table of Contents

Other manuals for BMC G2S A20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC G2S A20 and is the answer not in the manual?

BMC G2S A20 Specifications

General IconGeneral
BrandBMC
ModelG2S A20
CategoryMedical Equipment
LanguageEnglish

Related product manuals