G2S A20 / G2S C20 User Manual V1.0
34 / 39
23. EMC Requirements
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The
user of the device should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment -
Guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only for
its internal function. Therefore its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11
Class B
The device is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
/ flicker emissions
IEC 61000-3-3
Complies