19 Service Menu
NIPPY Clearway 2 Service Manual
2210.2657 Rev 3.0
3. Connect the Pressure measurement device.
4. Press ‘Start Blower [1] and Adjust P1 and P2 by pressing the
corresponding [-] and [+] buttons [2][3] until they match the pressure
measurement device reading. Press ‘STOP[4] to stop the blower.
Then press ‘Next’.
NOTE: The blower will default to creating a pressure of around 45
cmH
2
O, we recommended to calibrate this point. It is only required
to calibrate this one point.