Cleaning and maintenance
09/2011 1002410-04 Operating instructions – 35
5.2 Faults
Error codes are output to the display to simplify troubleshooting:
y If a fault occurs, the error code flashes on the display.
y If several errors occur, the error messages are displayed alternately.
y Error elimination is acknowledged by pressing the START/STOP button.
y If the error cannot be cleared, inform Customer Service - see page 2.
More detailed information regarding the error description and solutions is
available in 'Fehlercodes/Error Codes' (order no. 1008954-..).
5.3 Cleaning
DANGER!
Infections can be spread by skin contact!
All objects / device parts which may come into contact with
the user during use, must be disinfected after every use:
– Acrylic glass panel
– Handles and control panel
– Sensor and base station
– Protective goggles
– MP3 cable
– Adjustable air nozzles
NOTE:
Adhere to the stipulated times to ensure thorough disinfec-
tion. Follow the manufacturer's instructions for use.
5.3.1 Cleaning the surfaces
Acrylic glass surfaces
WARNING!
Do not rub with a dry cloth – danger of scratching!
Failure to comply voids any warranty claims.
– For the fast, hygienic and proper cleaning of acrylic
glass surfaces, use specially developed cleaning
agents.
– Concentrated disinfectants or solvents (e.g. Lysoform,
ethyl alcohol or other fluids containing alcohol) must