7.6.8 Regenerate dried out pH electrode
Introduction
If the glass membrane dries out it is important to restore the hydrated glass layer.
DescriptionStage
First try Method 1.1
If Method 1 does not help, try Method 2.2
If the electrode continues to be slow or not function properly, replace it with
a new one.
3
Regeneration methods
Use one or both of the following methods to regenerate a dried out pH electrode.
Method 1
Soak the electrode in 1 M KNO
3
for 5 hours or overnight.
Method 2
Soak the electrode in 0.1 M HCl overnight. Rinse it carefully with distilled water. Soak it
in 4 pH buffer for 1 hour.
ÄKTA avant User Manual 29035184 AE 311
7 Maintenance
7.6 Maintenance when required
7.6.8 Regenerate dried out pH electrode