9.10 Available Curves
Curves
The table lists data available in curves for ÄKTA avant.
DescriptionSam-
pling
frequen-
cy
RangeCurve
The first UV/Vis absorbance
signal of the U9-M monitor.
10 Hz-6000.000 to 6000.000
mAU
UV 1_280
The 2nd UV/Vis absorbance
signal of the U9-M monitor.
2 Hz-6000.000 to 6000.000
mAU
UV 2_0
The 3rd UV/Vis absorbance
signal of the U9-M monitor.
2 Hz-6000.000 to 6000.000
mAU
UV 3_0
The conductivity signal.5 Hz0.00 to 999.99 mS/cmCond
The conductivity signal as a
percentage of a set range.
1 Hz0.0% to 100.0%% Cond
The set concentration B or the
current value during a gradi-
ent.
1 Hz0.0% to 100.0%Conc B
The pH signal.1 Hz0.00pH
The set flow rate of the system
pumps.
1 Hz0.001 to 50.000
ml/min (ÄKTA avant
25)
System flow
0.01 to 300.00 ml/min
(ÄKTA avant 150)
The set flow velocity of the
system pumps. Only available
if a column is selected.
1 Hzcm/hSystem flow
linear
The system pressure signal (at
the system pumps).
10 Hz-1.00 to 20.00 MPa
(ÄKTA avant 25)
System pres-
sure
-1.00 to 5.00 MPa
(ÄKTA avant 150)
The temperature signal (in the
conductivity flow cell).
0.5 Hz0.0°C to 99.0°CCond temp
ÄKTA avant User Manual 29035184 AE 501
9 Reference information
9.10 Available Curves

Table of Contents

Other manuals for GE AKTA avant

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTA avant and is the answer not in the manual?

GE AKTA avant Specifications

General IconGeneral
BrandGE
ModelAKTA avant
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals