GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 4-7 - Keyboard Function Checks 4 - 25
4-7-1 Main Keyboard Assy Function Validation Procedure (cont’d)
6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the
scfpapitest program.
7.) Press the keyboard button and rotate rotary button. When you press or rotary keys, Log dialog box
displays the key log. Button name and displaying log name is like the table.
Figure 4-28 scfpapitest program
Table 4-7 Button and Rotaty Table
Button or Rotary Name The Name displayed in Log Button or Rotary Name The Name displayed in Log
Patient
New Patient Button
Active Mode Active Mode Button
Probe/Preset
Probe Button
CF CFM PushEncoder Button
Report
Report Button
PDI PDI Button
End Exam
End Exam Button
Dept (Button) Depth PushEncoder Button
Record
Record button
Dept (Rotary) Depth PushEncoder Rotary
P1
Print or record button1
PW PWD PushEncoder Button
P2
Print or record button2
CW CW Button
Reverse
Reverse Button
M M Mode PushEncoder
Userdefine 1
Userdefine 1 Button
Scan Area ScanArea Button
Userdefine 2
Userdefine 2 Button
M/D Cursor MD Cursor Button
Userdefine 3
Userdefine 3 Button
Zoom (Button) Zoom/Size PushEncoder Button
Userdefine 4
Userdefine 4 Button
Zoom (Rotary) Zoom/Size PushEncoder Rotary
Userdefine 5
Userdefine 5 Button
Measure Measure Button
L (Left)
Left Button
BodyPattern/Ellipse (Button)
BodyPattern/Ellipse PushEncoder
Button