EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P6

GE LOGIQ P6 Service Manual

GE LOGIQ P6
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #170 background imageLoading...
Page #170 background image
GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
5 - 40 Section 5-1 - Overview
Section 5-9Machine RFS
5-9-1 General
This describes general MRFS (Machine RFS) functionality.
1.) MRFS function is an automatic call for service from the system to the OLC when system detects
high temperature or high or low voltage.
5-9-2 Enabling MRFS
NOTE: In order to enable MRFS, System must be connected to network and should be checked out to
a product server. Default MRFS user should also be set.
1.) If a system is not already checked out, the system should be checked out to a product server. Refer
to 5-8-2) System Checkout Procedure
2.) MRFS function is an automatic call for service from the system to the OLC when system detects
high temperature or high or low voltage.
3.) Right Click on “Connect to GE” icon.
4.) When RFS window pops up, click on “Users” tab.
Figure 5-42 Contact GE Icon
Figure 5-43 User tab

Table of Contents

Other manuals for GE LOGIQ P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 and is the answer not in the manual?

GE LOGIQ P6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals