GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 4-3 - General Procedure 4 - 15
4-3-11 Backup and Restore Database, Preset Configurations and Images
NOTE: When upgrading from R1.0.X to R2.0.X, preset cannot be transferred.
4-3-11-1 Formatting Media
1.) To format the backup media, CD-R or DVD-R, select the UTILITY
key on the Front Panel.
2.) Select CONNECTIVITY
, then RERNOVABLE MEDIA. Properly label and Insert the backup
media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-13. It is best to use all capital letters with no
spaces or punctuation marks.
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-14. When the formatting
has been completed, press OK
to continue.
6.) If desired, verify that the format was successful by returning to
Utility>Connectivity>Rernovable Media and selecting VERIFY
as shown in Figure 4-11.
Figure 4-13 Format and Verify Media
Figure 4-14 Format Successful Pop-up Menu
Verify Format
Format
Media
Selection
Media Label

Table of Contents

Other manuals for GE LOGIQ P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 and is the answer not in the manual?

GE LOGIQ P6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals