PreparingthePatient
3.Applytheelectrodestothepreparedarea.
WARNING:
SHOCKHAZARDEnsurethatconductivepartsoftheelectrodesorleadwiresdonot
comeincontactwithotherconductiveparts.
Thiswouldcanceltheprotectionprovidedbytheisolatedsignalinput.
WARNING:
CONDUCTIVEPARTSKeeptheconductivepartsofleadelectrodesandassociatedparts
awayfromotherconductiveparts,includingearth.
4.Verifytheleadsareallconnectedandworkingproperly.
NOTE:
UseonlyelectrodesandcontactagentsrecommendedbyGEHealthcare.Thesignal
qualityonthelead-checkscreenwillnotbeindicateduntiltheRL/Nelectrodehas
beenapplied.WhenRA/Rbecomesdisconnected,thesystemwilldisplayLimbLead
Disconnected.
ApplyingRestingElectrodes
CAUTION:
PROPERLEADWIRECONNECTIONImproperconnectionwillcauseinaccuraciesintheECG.For
multi-link,traceeachindividualleadwireinthepatientcablemoduletothecoloredconnector
andthentotheproperelectrodetoensurethatitismatchedtothecorrectlabellocation
Standard12LeadPlacement
Toacquireastandard12leadECG,usetheplacementshowninthefollowingillustrations.
2047426-001KMACâ„¢60057