32 2049809-001 Rev B © 2010 by General Electric Company. All rights reserved.
Chapter 4: Installation and Setup
Ultrasound
Use only Corometrics 5700 Series ultrasound transducers with a Mini Telemetry System.
Remind the patient to use the headset / speaker to check for continual pickup of the fetal heart rate signal
following each fetal movement.
FECG
You may need to tape the transducer cable to the patient to prevent excessive tension on the legplate or
attachment pad.
Though it is not suggested to use FECG during ambulation, the recommended position for the legplate is on
top of the upper thigh instead of the inner thigh. This facilitates walking and minimizes uid contacting the
legplate.
Tocotransducer
Remember to place the receiver’s UA Mode Selector switch in the TOCO position.
When connecting or re-connecting a Corometrics Trimline Tocotransducer to the transmitter’s UA connector,
you must wait at least 10 seconds before pressing the monitor’s UA Reference button. If any device (monitor,
receiver, transmitter) is o, you must wait at least ten seconds from the time the last device is powered on.
IUPC
Remember to place the receiver’s UA Mode Selector switch in the IUP position.