34 2049809-001 Rev B © 2010 by General Electric Company. All rights reserved.
Chapter 5: Maintenance and Cleaning
The following table lists approved cleaning solutions.
Generic Formulation Maximum concentration level
Hydrogen peroxide 6 %
Sodium hypochlorite 100 parts per million (ppm)
CaviCide®
100 % spray (apply on equipment sprayed on cloth-
not directly on equipment.
Shoulder Strap Cleaning
1. Wipe any uids from the surface of the shoulder strap.
2. Dampen a soft cloth with cleaning solution and gently clean the strap.
3. Wipe with cotton cloth dipped in water to remove any traces of cleaning agent.
4. Dry with a soft, dry cloth.
The following table lists approved cleaning solutions.
Generic Formulation Maximum concentration level Expected life of the belt
Hydrogen peroxide 6 % Minimum 1000 cleanings
Sodium hypochlorite 100 parts per million (ppm) Minimum 1000 cleanings
CaviWipes™ Not applicable for CaviWipes™ Minimum 300 cleanings