Loading Base Image Software
Venue 50 – Basic Service Manual 8-15
5447566-100 English Rev. 7
Loading the System Software
NOTE: Before loading the system software, please ensure that the
power can be continuously supplied and there is no risk of
power cut off during loading procedure.
NOTE: Before starting this procedure, remove all probes and
peripherals.
NOTE: While it is believed to be unnecessary, it would not hurt to
disconnect the system from the network and remove all
transducers.
NOTE: Do not plug and unplug the probe, mount to or release from the
Docking Station/Cart during the upgrading process.
1. Insert the SD Card labeled “VENUE Software” into the SD
Card Socket.
2. Power on the system, then the software upgrading startup
screen displays. Time in the upper right corner shows
system time, and the time in the lower right corner shows
the time of upgrading process.