Setup reminders
Venue 50 – Basic Service Manual 3-3
5447566-100 English Rev. 7
Setup reminders
Average setup time
Unpacking the Venue 50: 20 minutes
Set up Venue 50 wo/options: 30 minutes (dependent on the
configuration that is required)
DICOM Network Configuration: 30 minutes
The Venue 50 installation and functional checkout will take
approximately one hour. Venue 50 console with optional
equipment may take slightly longer.
Setup warnings
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE
OF OPENING THE AC GROUND LINE (I.E., METER’S
GROUND SWITCH IS OPEN), DON’T TOUCH THE
ULTRASOUND SYSTEM!
CAUTION
To prevent electrical shock, connect the unit to a properly
grounded power outlet. Do not use a three to two prong
adapter. This defeats safety grounding.
CAUTION
Do not wear the ESD wrist strap when you work on live circuits
and more than 30 V peak is present.
CAUTION
Do not operate this unit unless all board covers and frame
panels are securely in place. System performance and cooling
require this.

Table of Contents

Other manuals for GE Venue 50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Venue 50 and is the answer not in the manual?

GE Venue 50 Specifications

General IconGeneral
BrandGE
ModelVenue 50
CategoryMedical Equipment
LanguageEnglish

Related product manuals