Using Vscan Air in home healthcare or pre-hospital emergency medical
Vscan AirUser Manual 5-83
Direction GP092020-1EN Rev 18
When time is critical and charging options are limited or unavailable
The following considerations should be made if the Vscan Air
system is used in home healthcare or pre-hospital emergency
environments where time is critical and the options to charge the
Vscan Air probe or the display devices are unavailable or
limited.
General considerations:
Make sure that both the Vscan Air probe and Vscan Air app
selected for use has been registered and are prepared as
described in section 3.
It is recommended for users to get comfortable with
handling the Vscan Air prior to taking the device into a home
healthcare or pre-hospital emergency environment.
To help uncover any unexpected challenges with a stable
connection between the Vscan Air probe and Vscan Air app,
the selected Vscan Air probe and the preferred display
device should be tested in an environment as close as
possible to the environment where the service is to be
provided.
If an internet connection is not available and//or a user
cannot sign in because of other reasons selecting Guest on
the Sign in page will give the access to scan and store
functionality. Review of images stored while using the guest
option is available.
Before taking the device for use in a pre-hospital or home
healthcare environment:
Check the display device charge level.
Check the Vscan Air probe battery charge level according to
the description in the “Vscan Air CL is OFF” section.
After each use in pre-hospital or home healthcare
environment:
Check the display device charge level.
Check the Vscan Air probe battery status according to the
description in the “Vscan Air CL is OFF” section.
If needed charge the Vscan Air probe to a 90-100% level as
described in the “How to charge the Vscan Air CL” section.
The Vscan Air probe will switch itself off when charging.
If needed charge the display device to a 90-100% charge
level.