9. MOUNTING AND INSTALLATION
These sensors are precision instruments with a resolution of 0.001mm.
The devices must therefore be handled with great care. This instruction
manual must be read and observed by every user.
9.1. Mounting for bending compensated tie bar strain measurement
9.1.1. Preparation
The sensors must always be placed in pairs onto a tie bar (1.1&1.2), (2.1&2.2) etc.
Position
The distance between the sensor and clamping
plate must be at least once the diameter of the tie bar so
that the deformation of the tie bar can be measured
evenly and correctly.
Surface
Oil and other surface contamination, or paint which
has been applied, must be removed at the site of
installation.
Sensor
Check the charge of the current batteries
(see Chapter 8 „Operation and control“)..
Check the foil on the sensor for signs of damage.