Servo-i Ventilator System Service procedures
Revision 02 Service Manual 5 - 5
5
Edit configuration
Available only in Standby mode.
Default Alarm limits can be changed in this window:
Adult alarm limits and Infant alarm limits
• Pressure – Upper alarm limit
• Minute Volume – Lower and Upper alarm limit
• Respiratory Rate – Lower and Upper alarm limit
• End Exp. Pressure – Lower alarm limit
• End Tidal CO
2
– Lower and Upper alarm limit
• Apnea time – Upper alarm limit.
Default configuration regarding 'General' and 'Units'
cannot be changed in this window. Access to the
Field Service System (FSS) is required.
Displayed values
The Measured value boxes on the User Interface
shows different parameters. Some of these
parameters are fixed and some are selectable in this
window.
NIV adult alarms and NIV infant alarms
• Pressure – Upper alarm limit
• Minute Volume – Lower and Upper alarm limit
• Respiratory Rate – Lower and Upper alarm limit
• End Exp. Pressure – Lower alarm limit
• End Tidal CO
2
– Lower and Upper alarm limit
The alarms listed above, except Pressure alarm, can
be permanently silenced in this window.
Default configuration regarding 'General' and 'Units'
cannot be changed in this window. Access to the
Field Service System (FSS) is required.
Copy configuration
Available only in Standby mode.
Default user configuration can be copied to / from a
PC Card.
Default configuration regarding 'General', 'Units' and
'Alarm limits' will be copied.
Useful when applying the same user configuration on
a number of units.
Set date and clock
Available only in Standby mode.
Date and time setting can be changed in this
window.
O
2
cell adaptation
Available only during ventilation.
Adjustment of the O
2
concentration reading.
Refer to the 'Servo-i Ventilator System – User's
Manual', chapter Operating your Servo-i / Operating
- Adjustments for further information.