9.12 Reference height / innage measurement
If the unit is tted with the additional load (see Figure 9-8) then 
reference height / innage measurement are possible.
•  Install the HERMetic equipment as per 9.9 “Installation of the 
instrument”.
•  Open the valve by turning the handle.
•   Put the tape cleaner on the “DOWN” position. Disengage 
the knob of the crank handle and lower the sensing probe 
into the tank by turning the reel. Make sure that the tape 
does not rub on any sharp edge when lowering as its 
insulation could be damaged.
•  When the sensing probe comes in contact with the dip/
datum plate record the distance shown on the reading index. 
See Figure 9-9. The exact distance from the plate to the 
reading index is (reading + 4 mm) or (reading + 5/32”) which 
is the reference height providing the reading index level has 
been adjusted to the zero ullage level of the tank. If the tank 
zero ullage is levelled above or below the reading index, an 
additional correction shall apply. For more details refer to 
section “Installation of the gauging system”.
•   Turn the tape cleaner on “UP” position
.
•  Switch on the unit and raise up the sensing probe until 
checking the oil/water interface if any (see details in section 
9.11 ”Ullage / interface measurement”). To get a better 
accuracy of the interface level, release the tape cleaner on 
the “DOWN” position during the nal checking. Calculate 
the free water height by subtracting the index reading to the 
reference height.
•   Reengage the tape cleaner on the “UP” position and raise 
up the sensing probe until checking the ullage (see details in 
section 9.11 “Ullage / interface measurement”). Release the 
tape cleaner for nal checking of the ullage. Calculate the 
innage by subtracting the index reading and the free water 
height to the reference height determined before.
•  When the measurements are completed, switch off the unit, 
engage the tape cleaner on the “UP” position and wind up 
the tape until the sensing probe is in the storage tube. The 
reading on the tape shall be less than 420 mm or 1 ft 5 inch.
•  Close the valve and disconnect the instrument from the nipple.
•  Put the end cap back on the valve.
Figure 9-8
+ 4 mm
(5/32")
=
+ 4 mm (5/32")
100
200
300
400
500
600
700
800
900
1000
1100
1200
0
Tape reading
Distance from dip/datum plate
to reading index
sensor 
with load
Dip/Datum plate
Figure 9-9

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell HERMetic UTImeter Gtex Chem and is the answer not in the manual?

Honeywell HERMetic UTImeter Gtex Chem Specifications

General IconGeneral
BrandHoneywell
ModelHERMetic UTImeter Gtex Chem
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals