37
10.2 Checking the battery
Please note that in case you have to change the
battery, it must be done only in a safe area. Refer to
section 10.3 “Battery replacement”.
10.2.1 Before starting gauging
Switch on the unit. The buzzer tones every 2 seconds if the 
battery is not too low.
The following sequences are displayed as per Figure 10-1, 
the 4
th
 sequence shows the remaining power of the battery  
in percentage and as a bar-graph.
If the power left is less than 50% we recommend to have 
a spare battery ready for exchange. See also 10.3 “Battery 
replacement”.
If the power left is less than 20% the message is blinking to 
advise that the power may not be enough to carry out all the 
work.
If the battery is too low, the unit will stop on the message 
“battery” as shown on Figure 10-2 and the buzzer tones 
continuously. Change the battery as per 10.3 “Battery 
replacement”.
If it is not possible to switch on the unit, the battery is out 
or work. Change the battery rst, as per 
10.3 “Battery 
replacement”.
ON
UTImeter
Battery
Init.
*********
25.94°C
Ver x.xx
97%
░ ░ ░ ░ ░
Figure 10-1
ON
UTImeter
Battery
Init.
*********
Battery
Ver x.xx
0%
Figure 10-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell HERMetic UTImeter Gtex Chem and is the answer not in the manual?

Honeywell HERMetic UTImeter Gtex Chem Specifications

General IconGeneral
BrandHoneywell
ModelHERMetic UTImeter Gtex Chem
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals