38 Honeywell Sensing and Control
Signal Conditioning, Self Calibrating Digital Indicators
Rev. G, 008-0608-00
Analog Output Incorrect
Make certain of the type of Analog Output (voltage or current)
that the channel is equipped with; see “Identifying the Output
Type” on page 36.
Use the DIAGNOSTICS -> DAC FULL SCALE and DIAGNOSTICS -> DAC
ZERO SCALE menu items to force the Analog Output to a known
output. Then, adjust your readout device, panel meter, PLC or
data acquisition system to match.
Check the output selection jumpers; see “Output Selection” on
page 36.
“APPLY 00000” on Power-up
The channel has detected that the transducer connected to the
instrument is different than the one the channel was last calibrat-
ed with. Because the CALIBRATION TYPE is set to either Shunt Cali-
bration or mV/V Calibration, the instrument is prompting you to
apply zero load in order to auto-calibrate to this new transducer.
Do one of the following, depending on the situation:
• Re-connecttheoriginaltransducertothechannelandre-
start the instrument.
• Press[ENTER]tore-calibratethechanneltothisnew
transducer using Shunt or mV/V Calibration and accept the
presently applied load as “0”. (In situations where one can’t
apply “0” load to an absolute pressure transducer or a load
cell with a pre-load, you can change the CALIBRATION DATA
-> ZERO-SCALE VALUE menu item from “0” to a load that can
be applied. For example, 14.7 PSIA or the known pre-load on
the load cell.
• UsetheCALIBRATE menu item to perform a Known-Load Cali-
bration with this new transducer.
Auxiliary Function Pins Not Operating
Make sure that the AUX1 FUNCTION or AUX2 FUNCTION SETUP menu
items are set correctly; if they are set to DISABLED then they will
not operate.
The Auxiliary Function (“AUX1” and “AUX2”) pins must be con-
nected to pin 10, not pin 8, to activate them.
Sensitivity to EMI/RFI
To obtain maximum immunity to electromagnetic or radio frequen-
cy interference, make certain that the shields of the transducer
cables are connected to the “cable shield connection screw” on
the rear panel of the instrument. See “Unamplified Transducer
Connection to Strain-Gage Input Channel” on page 27, “External
Arrangement of AC powered SC2000” on page 10 and “External
Arrangement of Model SC3004” on page 15.
Shunt Calibration Feature not Operating
To activate the channel’s shunt calibration function while the in-
strument is in RUN Mode, see “Shunt Calibration Check” on page
7. This serves as a quick check of the shunt cal feature.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell SC Series and is the answer not in the manual?

Honeywell SC Series Specifications

General IconGeneral
BrandHoneywell
ModelSC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals