hipot.com 30
Failure Mode Displays
Failure Mode Display* Test Type Description
AC Hipot Displayed if the leakage current exceeds the high limit
setting, but does not exceed the metering range.
AC Hipot Displayed if the leakage current does not exceed the low
limit setting.
AC Hipot Displayed if there is a short in the DUT during the rst
300 mS of test.
AC Hipot Displayed if the leakage current exceeds the metering
range and either a short or ashover has occurred
during the test.
AC/DC Hipot Displayed if the Continuity test within the hipot test fails.
DC Hipot Displayed if the leakage current exceeds the high limit
setting, but does not exceed the metering range.
Using the Display
Test Mode Displays
Test Mode Display Test Type Description
AC/DC Hipot & IR Displayed when the test voltage is ramping up from 0.0
VAC/VDC to full test voltage.
AC/DC Hipot Displayed when the test voltage has reached full potential.
Dwell is the amount of time the potential is held at the set
value.
AC/DC Hipot & IR Displayed when a test sequence has passed.
AC/DC Hipot & IR Displayed if the operator elects to stop a test in process.
This can be accomplished by pressing the RESET button
during the test.
IR Displayed while the test is in process. If the IR value exceeds
the set limits, the test will fail after the DELAY time has
expired.