Service
4.2.3Errorcodesanddiagnosticcodes
FlashcodeFault
Consequenceforthe
scooterComments
1
Batteriesmustbe
charged
ContinuestodriveThebatteriesaredischarged.Chargethebatteries
assoonaspossible.
2
BatteryvoltagetoolowStopsdrivingThebatteriesaredepleted.Chargebatteries.
Ifyouswitchthescooteroffforafewminutes,the
batteriescanoftenrecuperatetosuchastagethata
shortjourneyisstillpossible.Youshouldonlydothis
inanemergency,however,becausethiscausesthe
batteriestobecomeexcessivelydischarged.
3
BatteryvoltagetoohighStopsdrivingThebatteryvoltageistoohigh.Ifthebatterycharger
isconnected,disconnectitfromthescooter.
Theelectronicsystemchargesthebatterieswhen
runningdownhillandwhenbraking.Thisfaultis
causedwhenthebatteryvoltagebecomestoohigh
duringthisprocess.Switchthescooteroffandon
again.
4
PowertimeexceededStopsdrivingThemaximumcurrentwasexceededovertoolonga
period,probablybecausethemotorwasoverloaded
orhasbeenworkingagainstanimmovableresistance.
Switchthescooteroff,waitafewminutesandthen
switchonagain.
Theelectronicsystemhasdeterminedamotor
short-circuit.Checkthewiringharnessforshort-circuit
andcheckthemotor.
ContactyourInvacareprovider .
5
BrakefailureStopsdrivingEnsurethatthedisengagingleverisintheengaged
position.
Thereisadefectinthebrakingcoilorinthecabling.
Checkthemagneticbrakeandcablingforopen
orshort-circuitedcircuitry.ContactyourInvacare
provider .
6
Noneutralposition
whenswitchingscooter
on.
StopsdrivingDriveleverisnotinneutralwhenthekeyswitchwas
turned.Putthedriveleverinneutral,turnthepower
offandthenturnonagain.
Itmaybenecessarytoreplacethedrivelever .
ContactyourInvacareprovider .
7
Faultinspeed
potentiometer
StopsdrivingThedrivelevercontrolscouldbefaultyor
incorrectlyconnected.Checkthecablingforopen
orshort-circuitedcircuitry.
Potentiometerisnotcorrectlyadjustedandmustbe
replaced.ContactyourInvacareprovider .
8
MotorvoltageerrorStopsdrivingThemotororitscablingisdefective.Checkthe
cablingforopenorshort-circuitedcircuitry.
9
Miscellaneousinternal
fault
StopsdrivingContactyourInvacareprovider .
10
Push/freewheelmode
error
Stopsmoving
Thescooterhasexceededthepermissiblemaximum
speedduringpushingorfreewheeling.Switchthe
electronicssystemoffandonagain.
4.3Tighteningtorques
CAUTION!
Riskofdamagetomobilitydevicedueto
improperlytightenedscrews,nutsorplastic
connections.
Alwaystightenscrews,nutsetc.tothestated
tighteningtorque.
Onlytightenscrewsornutswhicharenotlisted
herengertight.
Thetighteningtorquesstatedinthefollowinglistarebased
onthethreaddiameterforthenutsandboltsforwhichno
specicvalueshavebeendetermined.Allvaluesassume
dryandde-greasedthreads.
ThreadTighteningtorquein
Nm±10%
M43Nm
M5
6Nm
1603798-B9