Channel Operation Commands
Copyright© Itech Electronics Co., Ltd. 31
Chapter10 Channel Operation Commands
INSTrument[:SELect]
This command is used to switch present channel.
Command syntax
INSTrument[:SELect] <cr>
Arguments
CH1|CH2|CH3
Query syntax
INSTrument[:SELect]?
INSTrument:NSELect
This command is used to switch present channel.
Command syntax
INSTrument:NSELect <nr>
Arguments
1-3
Query syntax
INSTrument:NSELect?
INSTrument:COUPle[:TRIGger]
This command is used to specify the channel affected by the trigger command.
If no channel is specified, default to current channel.
Command syntax
INSTrument:COUPle[:TRIGger] <cr1> <cr2> <cr3>
Arguments
CH1|CH2|CH3|NONE
Query syntax
INSTrument:COUPle[:TRIGger]?
INSTrument:COMbine:SERies
Set channel 1 and channel 2 of the power supply to serial state.