Load Functions
Copyright © ITECH Electronic Co., Ltd. 56
⚫ Pset=50.0W: constant power of CP Mode.
⚫ CF=1.414: crest factor. It cannot be edited.
⚫ PF=1.000: power factor, -1.000 to 1.000.
5. Press [Enter] to confirm.
6. Press [On] to enable load input. Loading will be started and the current will
sink.
4.3 DC Load Function
IT8600 electronic load can simulate the DC electronic load function according to
the setting. The user can select the present load function in the system menu.
When “Mode Select” is set as “DC”, the present load function of the electronic
load is in the DC mode.
IT8600 electronic load can be operated in the following four modes:
⚫ Constant current mode (CC)
⚫ Constant voltage mode (CV) (Only for IT8615/IT8615L)
⚫ Constant register mode (CR)
⚫ Constant power mode (CP)
4.3.1 Constant Current Mode (CC)
In this mode, the electronic load will sink a constant current in accordance with
the programmed value regardless of the input voltage. See figure 3-4.
Figure 3-4 CC mode