EasyManuals Logo
Home>Keithley>Measuring Instruments>2400

Keithley 2400 Service Manual

Keithley 2400
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
2400 SourceMeter
250V
PEAK
5V
PEAK
HI
LO
OUTPUT
250V
PEAK
250V
PEAK
EDIT
TOGGLE
POWER
RANGE
INPUT/
OUTPUT
4-WIRE
SENSE
DISPLAY
ON/OFF
TERMINALS
FRONT/
REAR
AUTO
RANGE
EXIT ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
230
1
67
89
4
+/-
5
STORE
RECALL
EDIT
Digital Multimeter
Input HI
Input LO
Model 2400
Fi
gure
1
-
1
Voltage verification front panel connections
2400-902-01 (G - Feb 2006)
1-9 Model 2400 Series User’s Manual
Performance Verification
5. Repeat the procedure for negative output voltages with the same magnitudes as those
listed in
Table 1-2.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-2
Output voltage accuracy limits
Model 2400
source range
Model 2400
output voltage setting
Output voltage limits
(1 year, 18•C–28•C)
200mV
2V
20V
200V
200.000mV
2.00000V
20.0000V
200.000V
199.360 to 200.640mV
1.99900 to 2.00100V
19.9936 to 20.0064V
199.936 to 200.064V