Table 2-3
Front panel voltage calibration
Source range* Source voltage Multimeter voltage reading**
0.2V
2V
20V
200V
+200.00mV
+000.00mV
-200.00mV
-000.00mV
+2.0000V
+0.0000V
-2.0000V
-0.0000V
+20.000V
+00.000V
-20.000V
-00.000V
+200.00V
+000.00V
-200.00V
-000.00V
___________mV
___________mV
___________mV
___________mV
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
____________Vm
**Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
2-11 Model 2400 Service Manual
Calibration
2400-902-01 (G - Feb 2006)