6-14 Source-Measure Concepts 2400 Series SourceMeter
®
User’s Manual
Operating boundaries
Source or sink
Depending on how it is programmed and what is connected to the output (load or source),
the SourceMeter can operate in any of the four quadrants. The four quadrants of operation
for the SourceMeter models are shown in Figure 6-4 through Figure 6-8. When operating
in the first (I) or third (III) quadrant, the SourceMeter is operating as a source (V and I
have the same polarity). As a source, the SourceMeter is delivering power to a load.
When operating in the second (II) or fourth (IV) quadrant, the SourceMeter is operating as
a sink (V and I have opposite polarity). As a sink, it is dissipating power rather than sourc-
ing it. An external source or an energy storage device, such as a capacitor or battery, can
force operation in the sink region. See Section 3, “Sink operation,” for more information.
Duty cycle
The duty cycle is the ratio of source on time to total time period expressed as a percentage.
For example, if the source is on for six out of 10 seconds, the duty cycle is 60%.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals