11-32 Triggering 2400 Series SourceMeter
®
User’s Manual
Figure 11-14
2430 Pulse Mode trigger model (remote operation)
Arm-In
Event
IMMediate
BUS
TIMer
MANual
TLINk
NSTest
PSTest
BSTest
Arm Event
Detector
Source Event
Detector
No
Yes
Idle
Another
Arm
?
Another
Trigger
?
Yes
No
Arm
Layer
Trigger
Layer
= GPIB Default
= Output Trigger
Trigger-In
Source
No
See Note
INITiate
?
Yes
SOURce
ARM
:DIRection
ACCeptor
ARM :SOURce
TRIGger
:DIRection
ACCeptor
TRIGger:INPut
SOURce
TRIGger:OUTPut*
SENSe
*GPIB default parameter for
TRIGger:OUTPut is NONE.
TRIGger:COUNt
<n>
1
ARM:OUTPut
NONE|TEXit
ARM:COUNt
<n>|INF
1
Note: The following commands
place the SourceMeter into
idle: DCL, SDC, ABORt,
*RST, SYSTem:PREset, and
*RCL.
TRIGger:SOURce
IMMediate
TLINk
SOURCE
Action
Pulse Delay
SOURce:PULSe:DELay
<n> 0.0 sec
SOURce:PULSe:WIDTh
<n>
0.0002
MEASURE
Action
Pulse Width
ARM:OUTPut
NONE|TENTer
Delay Event
Detector
Sense Event
Detector
Trigger-In
Source
TRIGger:INPut
DELay
Trigger-In
Source
TRIGger:INPut
SENSe
DELay
SOURce
TRIGger:OUTPut*
TRIGger:OUTPut*
SOURce
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals