18-10 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Table 18-5
ROUTe command summary
Command Description
Default
parameter SCPI
Source
memory
:ROUTe
:TERMinals <name>
:TERMinals?
Select in/out terminals: (FRONt or REAR).
Query in/out terminals.
FRONt ✓ ✓
Table 18-6
SENSe command summary
Command Description
Default
parameter SCPI
Source
memory
[:SENSe[1]] Sense 1 Subsystem: ✓
:DATA Path to SENSe[1] data. ✓
[:LATest?] Return only most recent reading.
:FUNCtion Select measurement function(s): ✓
:CONCurrent <b> Enable or disable ability to measure more
than one function simultaneously. When
disabled, volts function is enabled.
1
ON ✓✓
:CONCurrent? Query concurrent state. ✓
[:ON] <function list> Specify functions to enable
(VOLTage[:DC], CURRent[:DC], or
RESistance).
CURRent ✓✓
:ALL Enable all functions (concurrent
enabled) or enable ohms function
(concurrent disabled).
1
✓
:COUNt? Query number of functions that are
enabled.
✓
[:ON]? Returns list of functions that are enabled. ✓
:OFF <function list> Specify functions to disable:
(VOLTage[:DC], CURRent[:DC], or
RESistance).
VOLTage,
RESistance
✓✓
:ALL Disable all measurement functions. ✓
:COUNt? Query number of functions that are
disabled.
✓
:OFF? Returns list of functions that are disabled. ✓
:STATe? <name> Query state of specified function: 1 (on)
or 0 (off).
✓
1. For the Model 2430 Pulse Mode, concurrent measurements are always disabled. This command is invalid (error +831).