18-80 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Query :TRIGgered? Query triggered amplitude for fixed source
:TRIGgered? DEFault Query *RST default amplitude
:TRIGgered? MINimum Query lowest allowable amplitude
:TRIGgered? MAXimum Query highest allowable amplitude
Description This command performs the same as the [:IMMediate][:AMPLitude]
command except that the amplitude is not updated immediately.
With this command, the amplitude is updated when the SourceMeter is
triggered to perform a source-measure operation. For example, if the
instrument is waiting in the trigger layer for an external trigger, the
amplitude of the source will not update until that external trigger is
received by the SourceMeter. See Section 11, “Trigger models,” for
details on trigger model operation.
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or
MAXimum parameters on a lower source range will generate error -221
(Setting Conflict).
Set voltage limit
[:LEVel] <n>
:SOURce[1]:VOLTage:PROTection[:LEVel] <n> Set voltage limit for V-Source
Parameters 2400
<n> = -210 to 210 Specify V-Source limit
20 Set limit to 20V
40 Set limit to 40V
60 Set limit to 60V
80 Set limit to 80V
100 Set limit to 100V
120 Set limit to 120V
160 Set limit to 160V
161 to 210 Set limit to NONE
NONE Set limit to 210V
DEFault Set limit to 210V (NONE)
MINimum Set limit to 20V
MAXimum Set limit to 210V (NONE)