18-28 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
Varistor alpha
*RST
:SENS:FUNC:OFF:ALL
:SENS:FUNC:ON “VOLT”,“CURR”
:SOUR:FUNC:MODE CURR
:SOUR:CURR:STAR <n>;STOP <n>;MODE SWE
:SOUR:CURR:STEP <n>
:TRIG:COUN 2
:CALC:MATH:EXPR:NAME “VARALPHA”
:CALC:STAT ON
:OUTPUT ON
:INIT
:CALC:DATA?
Percent Deviation
NOTE Percent deviation is not a built-in math expression for remote operation. This
program example creates the following percent deviation (PER_DEV) calcula-
tion to test 10k
resistors. This user-defined math expression is added to the
catalog.
Where: RES is the actual measured resistance of the DUT.
10k is the reference value.
*RST
:SENS:FUNC:OFF:ALL
:SENS:FUNC:ON or “RES”
:CALC:MATH:UNIT “%”
:CALC:MATH:EXPR:NAME “PER_DEV”
:CALC:MATH:EXPR ((RES - 10e3) / 10e3) * 100
:CALC:MATH:EXPR:NAME “PER_DEV”
:CALC:STAT ON
:OUTPUT ON
:INIT
:CALC:DATA?
NOTE Parameter <n> referenced in the :SOUR:VOLT and :SOUR:CURR commands
above represent the actual numbers that the user would program. All other com-
mands should be entered as shown.
PER_DEV
RES 10k()
10k
------------------------------------
100×=
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 Series and is the answer not in the manual?

Keithley SourceMeter 2400 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals