36 TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011
8.2 Automatic adjustment using the internal weight
Factory setting in the models in the verifiable setting (ALJ/PLJ)
The automatic adjustment using the internal weight is started automatically:
when the scale was disconnected from the mains,
after the ON/OFF button is pressed in the stand-by mode,
after the temperature is changed by 1.5 degrees Celsjus for the non-
loaded scale plate / zero indication (this prevents adjustment when a
series of measurements is performed)
after 20 minutes, for the non-loaded scale plate / zero indication (this
prevents adjustment when a series of measurements is performed).
The automatic adjustment function is always on. However, it is always possible to
carry out manual adjustment (by pressing the CAL button) using the internal weight,
see chapter 8.3.
Automatic adjustment course:
The <Cal 25> symbol informs about the approaching adjustment.
In that time, the user should finish weighing.
The countdown starting from 25 s starts [CAL 25] [CAL 0].
During that period of 25 s, the adjustment can be stopped and delayed for 5 minutes,
pressing the ON/OFF button. As a result, the scale will be switched back to the
weighing mode, e.g. to complete the ongoing measurement.