1 Technical specification
KERN ALJ 160-4A ALJ 200-5DA
Product number / type TALJG 160-4-A TALJG 220-5-A
Weighing range (Max) 160 g 82 g/220 g
Interval (d) 0.1 mg 0.01 mg/0.1 mg
Reproducibility 0.1 mg 0.04 mg/0.1 mg
Linearity ±0.3 mg ±0.1 mg/0.2 mg
Settling time (standard) 4 s 10 s
Minimum part weight when
laboratory conditions*
1 mg 1 mg
Minimum part weight when
standard conditions**
10 mg 10 mg
Heating time 8 h
Adjustment weight internal
Number of reference pieces
when counting the number of
pieces
10, 25, 50, 100, selected freely
Weight units
ct, g, gn, lb, mo, oz, ozt,
tl (Hong Kong), tl (Singapore, Malaysia), tl (Taiwan), pen
Power supply
24 V DC, 1A
Operating temperature +15°C ... +30°C
Air humidity max. 80% (non-condensing)
Housing (W × D × H) mm 210 × 340 × 330
Wind breaker dimensions
(W × D × H) mm
160 × 140 × 205 (inner)
190 × 195 × 225 (outer)
160 × 170 × 225 (inner)
190 × 195 × 225 (outer)
Scale plate (stainless steel) Ø 80 mm
(Net) weight kg 6.5 kg 7 kg
Interface RS-232C
Contamination degree 2
Overvoltage category Category II
Installation height
above sea level
up to 4000 m
Location site Solely indoors