7.4 Weighing with tare
A tare value can be entered for the reference as well as for the bulk material
balance. Before setting a tare value select active balance,
see chap. 9.3.
7.4.1 Taring
Deposit weighing vessel. After successful stop check press the TARE button.
Zero display and the indicator [
] above NET appear.
The weight of the container is now internally saved.
Weigh the material, the net weight will be indicated.
After removing the weighing container, the weight of the weighing container
appears as negative display.
To delete the tare value, remove load from weighing pan and press the TARE-
key.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The
limit is reached when the total weighing range capacity is full.
7.4.2 Numeric entering of tare weight
Unload and reset to zero the balance.
Input known tare weight via the numeric keys with decimal point, and confirm
with TARE key. The entered weight will be stored as tare weight and displayed
with negative sign. The indicator [
] above NET appears.
Put the filled weighing container on the balance, the net weight will be
displayed.
The tare value remains stored until it is deleted with the TARE key.
The tare value will be rounded up according to the readability of the balance,
e.g. at a balance 60 kg max/5 g readability the input value of 103 g will be
displayed as -105 g.
7.4.3 Switch-over weighing unit
Depending on the model, pressing the UNIT key may switch over between g / kg
lb (only with menu adjustment F1 oFFUnit kg / lb).
The indicator [
] shows the active unit.