15 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value
of the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a monitoring of test resources, you can
improve this by means of linearization.
Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of weighing scales.
The adjustment weights to be used must be adapted to the weighing scale’s
specifications; see chapter 3.4 “testing instruments control”.
Provide for required adjustment weights; cf. subsequent table 1 or table 2.
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
After successful linearisation you will have to carry out calibration;
see chapter 3.4 “testing instruments control”.
Access to menu:
Switch-on balance and during the selftest press .
Use the numeric keys to input password „9999“ on display „Pin“
Confirm input by .
Table 1: Required adjustment weights KERN CFS
Max 1. 2. 3. 4.
300 g
50 g 100 g 200 g 300 g
3 kg
0.5 kg 1 kg 2 kg 3 kg
6 kg
2 kg 6 kg - -
15 kg
5 kg 15 kg - -
30 kg
10 kg 30 kg - -
50 kg
15 kg 30 kg 50 kg -