62 CFS/CCS-BA-e-1624
Ensure that there are no objects on the weighing pan.
Wait for stability display (indicator [
] atop
expires), then press .
Carefully place required adjustment weight into the center
of the weighing plate on display „LoAd“.
Wait for stability display, then press .
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will
return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
( ), repeat adjustment process.
14.2 Models CFS 6K0.1, CFS 15K0.2, CFS 30K0.5
Operation Display
Switch-on balance and during the selftest press .
„Pin“
Use the numeric keys to enter password:
Either input four times zero „0000“ as a standard
password, or input the user-defined password (input cf.
parameter „Pin“ chapter 13).
Confirm input by .
„Pin“ „----
When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried out on
both balances.
Select bulk material or reference balance via .
The appeared [
] displays the current balance.
Acknowledge with .
„tECH“ „LoCAL“
„tECH“ „rEmotE
Using
select the weighing unit [kg or lb] which will
be used for linearization. The appeared [
] displays the
current weighing unit.
Acknowledge with .
„tECH“ „Unit“