TEWJ-B-BA-e-2310 25
7.8.5 Gravitational constant place of location < grause >
Press and hold the TARE and ON/OFF buttons simulta-
neously to enter the setup menu.
Wait until the first menu item < Cal > is displayed.
Confirm by [ ] button, < CalExt > will be displayed.
Use the navigation keys to select < grause>.
Acknowledge using [ ] button, the current setting is dis-
played. The active digit is flashing.
Enter weight value and confirm using the [ ] button, nu-
merical input see chap. 3.2.2.
Weighing balance returns to menu.
Press repeatedly button to exit menu.

Table of Contents