33 PBS/PBJ-BA-e-1718
8.2.2 Adjustment test with internal weight
Condition: Menu setting „I-tESt“/ menu element 2
In weighing mode press . I-tESt is displayed.
If „i-tESt“ is not displayed, press to return into weighing mode and
activate menu element 2, see menu overview.
Press , checking is automatic.
Wait for a short period of time, the difference to the previous
adjustment is displayed.
Either
Press , the d-value is reset to zero.
Adjustment of balance by resetting.
or
Press , the d-value is not reset to zero. No
adjustment occurs.