63 PBS/PBJ-BA-e-1718
16.11 Animal weighing
(only possible with non verifiable setting)
This function is used for weighing animals. The animal symbol is displayed when
the animal weighing mode is active.
Select the menu element 53 in the menu overview.
Place the weighing container on the scale pan and press the button.
Note:
When the weighing container is placed on the scale pan, data output is possible. It is
not a fault. This no malfunction.
Place an animal on the scale pan. Its weight must exceed 50 times the zero
range.
The value will be given automatically when the weighing value is stabilised.
Press the button or remove an animal from the scale pan.
If the displayed value is stable, but it lies below 10 times of the zero range,
the balance will be zeroed automatically. Any remaining materials on the
scale pan (excrement or skin) will be automatically written off and zeroed. If
the balance is not zeroed, the zero range value must be increased (refer to
chapter 10.7).
Notes:
The stand-by mode is not provided for the animal weighing function.
Press the button to switch over to the mains supply stand-by mode.
When live animals are weighed in the animal weighing mode, automatic
extension of stability detection band is carried out. Repeatability of measuring
data is a bit lower than in other operational modes.
If an animal being weighed cannot be controlled and the automatic printing
function does not respond, you can press the button to output the display
value. Then remove an animal from the balance. Even if the stability symbol
appears before an animal is removed from the balance, the data will not be
printed again.
The stability symbol will appear earlier due to setting the wider stability
detection band in the menu.
If the balance returns to the zero point too slowly, the zero range is to be set to
a higher value.
The initial taring function (chapter 11.2.1) cannot be used together with the
animal weighing function.
TARE /
TARE /
ESC
ON/OFF
PRINT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PBJ 620-3NM and is the answer not in the manual?

KERN PBJ 620-3NM Specifications

General IconGeneral
BrandKERN
ModelPBJ 620-3NM
CategoryScales
LanguageEnglish

Related product manuals