Insert new module for the Autolab 8-Series
13 | Page
Figure 12 An additional SMB cable must be placed between the ECD and the control amplifier
The ECD installation can be tested using Nova (version 1.3 or higher), as explained in
Section 14 of this document.
The ECD module installed can also be tested with the GPES software following the
instructions provided in the Autolab 4.9 Installation and Diagnostics manual.
8 Analog bus adjustments for the ADC10M module
For the ADC10M module, a ferrite bead should be placed in the Analog bus (see Figure
13). For instruments with serial numbers AUT83340 the ferrite bead is already installed
and this modification is not required.