Insert new module for the Autolab 8-Series
14 | Page
Figure 13 A ferrite bead must be installed on the analog bus connection to the ADC10M module
Follow these steps to install the ferrite bead, supplied with the module:
1. Locate the ADC10M wire 3 (see Figure 13).
2. Unscrew the wire at the ADC10M end of the Analog bus and slide the wire
through the ferrite bead. Reattach the wire end to the module connector.
3. Repeat step 2 for the two other wires.
The ADC10M installation can be tested using Nova (version 1.2 or higher), as explained in
Section 14 of this document.
Note: the ADC10M is NOT supported in the GPES software.