Technical details
179
Imperial Metric
Height with machine lid
Height without machine lid
32 7/8"
32 5/16"
835mm
820mm
Width
35 3/8" 898mm
Depth
Depth with door open
mm
mm
Wash cabinet dimensions:
Height
Width
Depth of upper basket/lower basket
20 9/16"
21 1/8"
20 9/16"
520mm
530mm
474mm/520mm
Weight (net)
216 lbs 98kg
Max. load capacity of open door
81.6 lbs 37kg
Voltage, rated load, fuse rating See data tag See data tag
Power cord
Approx. 5' 9" ft. Approx. 1.8m
Water connection water temperature:
Cold water/steam condenser
Hot water/DI water (optional)
max. 68 °F
max. 150 °F
Max. 20°C
Max. 65°C
Static water pressure
max. 145 psi Max. 1,000kPa pressure
Minimum water connection flow pressure:
Cold water/steam condenser
Hot water
DI water (optional)
14.5 psi
5.8 psi
4.4 psi
100kPa pressure
40kPa pressure
30kPa pressure
Recommended water connection flow pressure:
Cold water/hot water
DI water (optional)
Steam condenser
29 psi
29 psi
14.5 psi
≥200kPa pressure
≥200kPa pressure
≥100kPa pressure
DI water supply without pressure (optional)
1.3-8.7psi 8.5-60kPa
Drain height
min. 11 3/4" ft, max. 3' 3" ft Min. 0.3m, max. 1.0m
Drain hose length
max. 13' 1" ft Max. 4.0m
Operation (according to IEC/EN61010-1):
Ambient temperature
Relative humidity maximum
linear decrease to
Relative humidity minimum
40 °F to 104 °F
80% for temperatures up to 88°F
50% for temperatures up to 104°F
10%
5°C to 40°C
80% for temperatures up to 31°C
50% for temperatures up to 40°C
10%
Storage and transportation conditions:
Ambient temperature
Relative humidity
Air pressure
- 4°F to 140°F
10% to 85%
7.25psi to 15.37psi
- 20°C to 60°C
10% to 85%
500hPa to 1060hPa
Altitude above sea level (according to
IEC/EN61010-1)
up to 6,561 ft* Up to 2,000m*
Type of protection (according to IEC60529)
IP20
Soiling level (according to IEC/EN61010-1)
2
Overvoltage category (according to IEC60664)
II
Sound emission values in dB (A),
sound pressure level LpA during cleaning and dry-
ing phases
<70
Certifications CAN/CSA-C22.2 No. 61010-1-04, CAN/CSA-C22.2 No. 61010-2-040, UL
Std. No. 61010-1 (2nd Edition), IEC 61010-2-040:2006
Manufacturer’s address Miele & Cie. KG, Carl-Miele-Strasse 29,
33332 Gütersloh, Germany
* If installed above 6,561ft (2,000m), the boiling point of the wash water will be lower. In this case, the disinfecting tempera-
ture and the holding time might need to be adjusted.