A - 7
WARNING
Do perform all network functions of data communication within an enclosed network.
A.6.2 Cellular Specifications
Test conditions Meets the following conditions simultaneously:
Number of the equipments supported by a single AP: ≤ 3
Each equipment can communicate with the CMS.
The weakest strength of the AP signal where the equipment is located is
not less than -65 dBm.
The distance between the interfering devices and the equipment is
greater than 20 cm. A Wi-Fi interference (no greater than -85 dBm) in the
same channel and a Wi-Fi interference (no greater than -50 dBm) in an
adjacent-channel are presented synchronously. The interfering devices
include, but are not limited to, 2.4 G wireless devices, cellular mobile
networks, microwave ovens, interphones, cordless phones, and ESU
equipment. The interfering devices do not include Wi-Fi devices.
Network type 4G, 5G
Operating frequency 4G module (EU):
LTE FDD: B1/B3/B7/B8/B20/B28A
LTE TDD: B38/B40/B41
WCDMA: B1/B8
GSM: B3/B8
5G module:
NR: n1/n2/n3/n5/n7/n8/n12/n20/n28/n38/n40/n41/n48/n66/
n71/n77/n78/n79
LTE-FDD: B1/B2/B3/B4/B5/B7/B8/B9/B12/B13/B14/B17/B18/
B19/B20/B25/B26/B28/B29/B30/B32/B66/B71
LTE-TDD: B34/B38/39/B40/B41/B42/B48
Standard/Modulation mode LTE, 5G Sub-6GHz
Alarm delay time of network
disconnection
The message and alarm delay of network disconnection between the
equipment and CMS: ≤ 14 s.
Pre-hospital network performance
System capacity, interference immunity
and network stability
Meets the following requirements:
The total delay of data transmission from the equipment to the CMS: ≤
10 s.
The data loss percentage of cellular communication over a 4-hour
period: ≤ 0.5%.
Test conditions Meets the following conditions simultaneously:
With a distance greater than 2m from each other, three equipments
operate properly at the same time.
Each equipment can communicate with the CMS.
The weakest signal strength where the equipment is located is greater
than -95 dBm.
The distance between the interfering devices and the equipment is
greater than 20 cm. The interfering devices include, but are not limited
to, 2.4 G wireless devices, cellular mobile networks, interphones, cordless
phones.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D30 and is the answer not in the manual?

Mindray BeneHeart D30 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D30
CategoryMedical Equipment
LanguageEnglish

Related product manuals