P/N: 046-003757-00 Rev A
Quick Reference Guide
13
e. Select Adjust to adjust the PAWP reference line to the desired position
f. Select Confirm to enter the PAWP measurement
g. Select
1. Next Measurement to start a new measurement as needed
2. Exit to return to normal screen
h. Select Main Menu
(quick key) or to return to the main display
or select Calc>> to access the Hemodynamic Calculations menu
Note: PA alarms are inactive when a PAWP measurement is initiated.
Note: Follow manufacturer’s suggested procedures and hospital policy for
PAWP balloon inflation.
Initiate Hemodynamic Calculations
Hemodynamic calculations are a set of values that are used to determine the
hemodynamic status of the patient.
To initiate hemodynamic calculations on the DPM 6 or DPM 7 bedside
monitor:
a. Select CALC>> from the C.O. menu or PAWP menu or HEMODYNAMIC
from CALCULATIONS menu (
or Main Menu quick keys)
b. HR, Art (Mean), PA (Mean), CVP, and average CO measurements are
automatically entered from the current values
c. Select remaining parameter and manually enter
d. Select Calculate to complete calculation
e. Select Record to print the currently displayed Hemodynamic Calculations
f. Select Main Menu
(quick key) or to return to the main display
Note: Hemodynamic Calculations are printed at the DPM 6 and DPM 7 bedside
monitor only.