To adjust a patient’s arrhythmia threshold setting on the V12 or V21 bedside
monitor:
a. Select ECG waveform to enter ECG Setup tab
b. Select Arrhy Setup tab
c. Select individual arrhythmia threshold
d. Scroll and select value or enter value using on screen keypad
e. Select Accept to save settings and return to the main screen
Or
a. Select Alarms tab from Navigation Area
b. Select Arrhythmia to enter the Arrhythmia tab
c. Select Setup
d. Select individual arrhythmia threshold
e. Scroll and select value or enter value using on screen keypad
f. Select Accept to save settings and return to the main screen
RESET ALARMS (Version 2.5 or greater)
Select Alarm Acknowledge in the Navigation Area on the V12 or V21 bedside
monitor to silence all current alarms for a preconfigured time interval or until
the alarm condition is resolved.
Panorama® Central Station interface: Selecting Alarm Acknowledge on the
V12 or V21 bedside monitor will silence an alarm on the Panorama Central
Station, WorkStation, and ViewStation for a preconfigured amount of time.
Selecting Mute at the Panorama Central Station will not silence an alarm at the
V12 or V21 bedside monitor.
PAUSE ALL ALARMS (Version 2.5 or greater)
Select Audio Pause in the Navigation area on the V12 or V21 bedside monitor
to silence all current and future alarms for a preconfigured time interval.
Panorama® Central Station interface: Selecting Audio Pause on the V12
or V21 bedside monitor will pause active alarms on the Panorama Central
Station, WorkStation, and ViewStation for a preconfigured amount of time.
Selecting Mute at the Panorama Central Station will not pause alarms on the
V12 or V21 bedside monitor.