User and Service Manual XLTEK EEG32U Amplifier
17
To ensure the validity of signals, do not operate the device near any sources of electromagnetic interference
.
Turn off the system power before cleaning. Prevent detergent solution or cold sterilization agents from seeping
into the electronics of the system. Be careful around all connectors and edges. Do not use abrasive agents.
The accessories of this device may include several kinds of disposable, sterile needle electrodes. These needles
may be labeled as STERILE. The EEG32U amplifier is not sterile and cannot be made sterile.
The sale, distribution, or use of this device is restricted to, by, or on order of a physician.
Third-party software installed on the acquisition computer may interfere with the operation of the Natus
software. Please consult Natus Technical Support before installing third-party software on the computer.
This device does not contain any hazardous components; therefore, no special precautions are required for their
disposal.
No modification of this equipment is allowed.
Electrical Warnings and Cautions
XLTEK systems are intended for connection to a properly grounded electrical outlet only.
The EEG32U is an electrically isolated, electrically floating device. The computer that is part of the system
should be attached to a properly grounded electrical outlet.
Conductive parts of electrodes and their connectors are not to contact other conductive parts and earth.
Do NOT connect electrode inputs to earth ground. The EEG32U amplifier contains warning symbols reminding
you that the connections are intended for isolated patient connections only. Connecting an earth ground
might result in electrocution.
Do NOT turn on the system power until all cables have been connected, verified and visually inspected for any
damage. Failure to inspect the cables may result in electrocution. Verification of electrical safety should be
performed routinely.